Pall AcroPak Inline Sterile Capsule Filter, 0.8um /0.2um, Pall 12203
search
  • Pall AcroPak Inline Sterile Capsule Filter, 0.8um /0.2um, Pall 12203

Pall AcroPak Inline Sterile Capsule Filter, 0.8um /0.2um, Pall 12203

$77.00

Cyriva Pall AcroPak STERILE inline capsule filter, sold as a pack of 3 filters.

Polyethersulfone filter membrane with a 0.8um pore prefilter followed by a 0.2um final filter pore size.

Sterilized by gamma radiation

Pall catalog 12203

Quantity

  Security policy

We do not sell customer sales data.

  Delivery policy

Orders are generally shipped within 1 business day after receipt of payment, Monday through Friday.

  Return policy

Sterile items cannot be returned. All other items must receive a return authorization in writing.

The Cytiva - Pall AcroPak capsule filter is made with a Supor membrane filter, yielding high flow rates and throughputs, and is ideal for solutions where low protein binding is required. Not recommended for use with some ketones.
Efficiently processes up to 2 L of chemical and biological fluids.
Sterilization by gamma irradiation eliminates potential contamination by EtO residuals.
Upstream vent prevents vapor lock.
Built-in prefilter extends filter life when particulate-laden solutions such as serum-containing media are processed.

Materials of Construction

Filter Media: Supor [hydrophilic polyethersulfone (PES)] membrane
Housing, Vent Plug: Polypropylene
Sealing Technology: Thermal bonding
Filling Bell: Polycarbonate

Pore Size

0.8/0.2 μm

Effective Filtration Area

20 cm2

Dimensions

Housing Length: 8.3 cm (3.3 in.)
Housing Diameter: 6.7 cm (2.7 in.)

Inlet/Outlet Connections

6.4 - 12.7 mm (1/4 - 1/2 in.) diameter stepped hose barb with female luer slip interior and filling bell outlet

Typical Hold-Up Volume (with 10 psi air purge)

≤ 2.5 mL

Maximum Operating Temperature and Pressure

60°C (140°F) at 1.0 bar (100 kPa, 15 psi)
4.1 bar (410 kPa, 60 psi) at 21 - 24°C (70 - 75°F)

Recommended Integrity Test Minimum Bubble Point - Water

3.5 bar (350 kPa, 51 psi)

Typical Water Flow Rate

mL/min at 0.1 bar (10 kPa): 40 mL/min/psi: 28

Bacterial Retention

Lot samples retain a minimum of 107 cfu/cm2 of B. diminuta per modified ASTM F838, current revision

Endotoxin Level

< 0.25 EU/mL using Limulus Amoebocyte Lysate (LAL) test

Biological Safety

Passes United States Pharmacopeia (USP) Biological Reactivity Test, In Vivo <88>

Sterilization

Can withstand one autoclave cycle at 125°C (257°F) for 60 minutes; water wet capsule prior to autoclaving
PN 12203 is sterilized by gamma irradiation and individually packaged

Applications

Designed to add convenience to small- and medium-volume filtration (sterile and non-sterile)
Suitable for use with fluids containing dilute proteins, preservatives, or other critical components
Sterile filtration of media and buffers

3 filters per pack

CYTIVA Pall Catalog 12203 

12203

Specific References

MPN
12203
New
Comments (0)
No customer reviews for the moment.